Practical example: Import GC/MS measurement values to your LIMS system

In the following practical example, we show you how to connect common MS GC/MS analysis devices (gas chromatography/mass spectrometry) e.g. from Thermo Fisher directly to your LIMS via interface and automatically import measurement values.

Every analysis device that generates an export file with measurement values can be connected to LDB. To connect your existing GC/MS analysis device to your LDB LIMS, you need an export file and access to the LDB LIMS import interfaces. You can generate the export file directly in the GC/MS. The LDB import interfaces can be found under Menu bar -> Samples -> Import Interfaces.

To connect your GC/MS, perform the following six steps:

  1. First, review the export file. In the following
    In Excel:

In a text editor:

The following points are noteworthy:

  • There is no sample number for automatic assignment
  • The attribute row is in row 8
  • TabStop is used as a delimiter
  • Not all rows have a result
  • For not all listed parameters, a peak / concentration was determined by the GC
  1. Create a new import interface.
    Specify the properties of the file to be processed

  2. Upload the file as a sample file

  3. In the import mapping, define which information from the file is assigned to which location in the Labordatenbank

  4. Upload the measurement file of your GC for a sample

  5. Check the import

Laatste wijziging: 21.08.2026

Algemeen

Leveranciersbeoordeling

Materiaal

Medewerkers

Klanten

Opdrachten

Monsters

PDF-sjablonen

Bijlagen

Recepten

Verslagen

Editor voor rapporttabellen

Interfaces

Kompetenzen

AI-functies

Inleidende fase

Analyses

Sjablonen

Klantenportaal (optioneel)

Aanbiedingen

Facturen

Parameters

Rekenwerk met parameters

Webservices

Transformatiecode

Controleplannen / Grenswaarden / Specificaties

Documentbeheer

Vragen en antwoorden

Testmiddelen

Opleidingen voor medewerkers

8D-rapport

Overige

Bestandsbeheer

Processen